DX-100R Automatic Hematology Analyzer
This high-precision automatic blood analyzer applies advanced nucleic acid fluorescent staining and latex-enhanced immunoturbidimetric technology. Designed to assist with early screening and diagnosis, it is suitable for identifying conditions such as leukemia, malaria, infections, and other blood-related disorders.
Its modular design, intelligent fluorescence analysis capabilities, and compatibility with various sample types make it a reliable solution for modern clinical laboratories
Main Features
The DX-100R is a sophisticated automatic hematology analyzer that utilizes advanced nucleic acid fluorescent staining and latex-enhanced immunoturbidimetric technology. It supports early screening and diagnosis of conditions such as leukemia, malaria, infections, and a range of blood-related disorders. Designed to meet international standards, it is ideal for laboratories seeking accurate, high-throughput results with advanced diagnostic capabilities.
| Measurement Principle |
5DIFF: Flow cytometry with nucleic acid fluorescent staining CBC: Sheath flow impedance RET: Flow cytometry PLT-F: Flow cytometry WPC: Flow cytometry |
|---|---|
| Parameters |
Blood sample: report parameters ≥36 items; Blood sample: research parameters ≥29 items; Body fluid sample: report parameters ≥7 items; Body fluid sample: research parameters ≥10 items; Blood sample: 3 scattergrams (WBC, DIFF, RET), 2 histograms (RBC, PLT); Body fluid sample: 1 scattergram (DIFF), 1 histogram (RBC) |
| F-PLT Channel | None |
| WPC Channel | None |
| Throughput |
CBC+DIFF: 100 T/H CBC+DIFF+RET: 83 T/H Body fluid mode: 17 T/H |
| Testing Modes | CBC, CBC+DIFF, CBC+DIFF+RET, CBC+RET |
| Sample Type | Venous whole blood, pre-dilution blood, body fluid |
| Sample Aspiration Volume |
Whole blood mode: ≤88 μL Pre-dilution mode: ≤70 μL Body fluid mode: ≤200 μL |
| Sample Loading | Automatic / Manual |
| Linear Range |
WBC: 0–500×10⁹/L RBC: 0–8.6×10¹²/L HGB: 0–260 g/L PLT: 0–5000×10⁹/L |
| Reagents |
Diluent (DXS) Lysing reagents: 10-DHL, 10-DDL, 10-DWL Fluorescent dyes: 10-DWF, 10-DDF, 10-DRF Additional diluent: 10-DRS Cleaning Solution: DCC |
| Language | Chinese, English |
| Loader Capacity | 5 racks, 10 samples per rack |
| LIS Transfer | Bidirectional LIS supported |
| Retest Function | Automatic retest at local or other machines if results are abnormal |
| Powerplant | External air-compressor |
| QC Mode | L-J, X-Bar, X-BarM floating mean QC |
| Environmental Conditions |
Temperature: 15℃–30℃ Relative Humidity: 30%–85% Atmospheric Pressure: 70–106 KPa |
| Power Supply | 200V–240V~; 50/60Hz |
| Size (mm) | Width 675 × Depth 785 × Height 885 |
| Weight | ≤85 Kg |