CM-800 Clinical Chemistry Analyzer

The CM-800 is a high-speed clinical chemistry analyzer that can process up to 800 tests per hour. It features a dry heating system for temperature control, a precise sample handling system, and smart online loading. This combination ensures faster results, improved accuracy, and a smoother, more efficient user experience.

Main Features

High-Efficiency Sample Processing

  • Consistent throughput of 800 tests/hour (T/H)
  • Optional ISE module with 400 T/H capacity
  • 137 sample positions and 114 reagent positions for continuous workflow

Dry Heating Thermostatic System

  • Direct dry heating ensures quicker warm-up and reduces pre-heating time
  • Maintains stable reaction temperature at (37.0 ± 0.2)°C in cuvettes

High-Efficiency Triple Mix Bars

  • Triple mixing bars with dual-stage cleaning to prevent contamination
  • Real-time stirring speed monitoring for reliable performance

Accurate Sampling System

  • Polished sampling probes (internal & external) for high precision
  • Built-in liquid volume tracking and level detection to ensure accuracy

Intelligent Online Loading System

  • Reagents can be added during testing without stopping the machine
  • Peltier cooling system extends maintenance intervals and lowers repair risk

9-Step Warm Liquid Washing

  • 360° probe cleaning with cross-contamination rate below 0.05%
  • Anti-overflow design prevents liquid spillage during washing

Technical Specifications

  • Test Principles: Latex immunoturbidimetry, immunoturbidimetry, colorimetry
  • Test Methods: Rate method, end point method, fixed-time method
  • Dimensions: 1640 mm (L) × 940 mm (W) × 1215 mm (H)
  • Gross Weight: 500 kg
  • Sample Positions: 137 positions
  • Reagent Positions: 114 positions
  • Sample Types: Serum, plasma, whole blood, urine, and cerebrospinal fluid
  • Application Scenarios: Hospital Labs, Diagnostic Centers, Academic Research Institutes
  • Water Consumption: No more than 40 L purified water per hour
  • Dilution: Automatic pre-dilution
  • Minimum Reaction Volume: 100 μL
  • Minimum Sampling Volume: 1.5 μL